Barcode:
857268006080
Description
The Joerns DermaFloat APL mattress, code DFAPL-3680-M, is the mattress only. It is 80 in long and 8 in deep, rated 350 lbs (159 kg), and needs the DermaFloat APL control unit, sold separately.
What does it do and how does it work?
- Twenty cells alternate in adjacent pairs, while integrated low air loss moves moisture away under the therapy pad.
- Three cycle times are available: 5, 10 or 15 minutes, with eight comfort settings.
- Autofirm inflates the surface for a transfer or care task; Therapy holds it steady.
- Fowler Boost adds air at the sacrum when the head of the bed is raised 30 degrees or more.
- The panel locks automatically; an audible power-failure alarm has a reset.
- A CPR valve deflates the surface quickly.
Who is it for?
- Joerns describes the APL for treatment and prevention of pressure ulcers up to stage IV; a clinician decides whether it fits the person.
- It is not recommended for an unstable spinal fracture without a doctor's agreement.
- This code is the mattress only. Order the DFAPL-CU control unit if the bed does not already have one.
Will it fit the bed?
- 80 in long and 8 in deep (203 by 20 cm). We confirm the width before you order.
- 350 lbs (159 kg) is the mattress limit; the bed frame's own safe working load applies if lower.
- Ten straps secure the mattress to the moving part of the frame.
- Check the rail gaps. Health Canada recommends under 120 mm (4 3/4 in) within and under the rail, and under 60 mm (2 3/8 in) at the rail ends. See bed rails.
- A used mattress cannot be returned. Measure first and read the return policy.
Power, oxygen and the cover
- Cells stay inflated about 12 hours in a power cut if the control unit remains connected.
- Use oxygen only with a nasal cannula, mask or half-length tent; a full tent past the mattress is a fire risk.
- The therapy pad is low shear and low friction, treated with an antibacterial and antifungal agent, fluid proof and vapour permeable.
- Wipe the cover clean; do not steam clean, and no machine-wash cycle is published.
- CairRails side air bolsters are part of the system; they do not guarantee against a fall or entrapment.
How does it compare?
- DermaFloat LAL: a 10 in replacement rated 500 lbs with a lower safety mattress.
- DermaFloat APL control unit: the pump this mattress needs.
- Alternating pressure mattresses: the wider category.
Frequently asked questions
Do I need the control unit?
Yes. This code is the mattress only; the pump is the separate DFAPL-CU control unit, and the complete system is a third code.
How long does Autofirm stay on?
Autofirm inflates the surface for a transfer or care task, then returns after about 12 minutes.
Does the mattress roll up?
Yes. The mattress rolls up for storage and for moving it between rooms.
Can the cover be machine washed?
No. Wipe it clean and do not steam clean.
Funding may help: see who may help pay. Warranty terms are in the Warranty tab.
Specifications
- Weight capacity
- 350 lbs (159 kg) on the mattress
- Product weight
- 22 lbs (10 kg)
- Construction
- 20 individual 8 in air cells, nylon with polyurethane
- Cover
- Low shear, low friction therapy pad treated with an antibacterial and antifungal agent; fluid proof and moisture vapour permeable
- Cover care
- Wipe clean; do not steam clean
- Therapy modes
- Autofirm, Therapy (static) and Alternating
- Cycle time
- 5, 10 or 15 minutes
- Comfort settings
- Eight settings
- Alarms
- Audible power-failure alarm with reset
- CPR release
- CPR quick deflation
- Airflow
- Up to 40 litres per minute in static therapy or after a cycle
- Power failure protection
- Cells stay inflated about 12 hours if the control unit stays connected
- Controls
- Automatic panel lockout
- Mattress length
- 80 in (203 cm)
- Mattress height
- 8 in (20 cm)
- Maker item
- DFAPL-3680-M (mattress only)
- Straps
- Ten straps to the moving part of the bed frame
- Fowler Boost
- Inflates the sacral section when the head of the bed is at 30 degrees or more
- Control unit
- DFAPL-CU, sold separately
- Warranty
- 2 years mattress; 1 year cover and electromechanical parts
- Product type
- Alternating pressure and low air loss mattress
Figures from the manufacturer. Options and seat sizes can change them; we confirm yours before ordering.
Documents
Delivery and service
Delivery and setup
We deliver and set up equipment across Toronto and the GTA, and show you how everything works. Custom and special-order products can take longer; we confirm timing with you before delivery.
Service and repairs
Our technicians service what we sell, at home or in our shop.
Try before you buy
Visit a showroom in Etobicoke or Brampton to try products first, or book a free home assessment.
Warranty
Warranty details
Joerns Healthcare warrants the DermaFloat APL support surface (mattress and control unit) against defects in workmanship and materials under normal and proper use. The warranty covers the original purchaser.
- Mattress: 2 years.
- Cover and electromechanical parts (compressors, valves, circuit boards, hoses and couplers): 1 year.
- Joerns supplies replacement parts, or services or replaces a defective product, at its option.
- Excludes burns, improper chemicals, needle punctures, cuts or abrasions, excessive loads, staining, improper maintenance or cleaning, use against the manual, and unauthorized modification or repair.
Questions about this product?