Description
The Span America PressureGuard Span-Care Convertible, model SC803629, is an 80 x 36 x 7 in (203 x 91 x 18 cm) air and foam mattress for a hospital bed.
What does it do and how does it work?
It spreads pressure with four interconnected air cylinders inside a foam shell.
- The cylinders have pressure relief valves, so the surface self-adjusts.
- The Geo-Matt foam top has more than 800 cells and a Heel Slope.
- Safety Edge bolsters around the perimeter add support at the edge.
- The LifeSpan cover has Shear Transfer Zones under the heels, sacrum and scapula.
Who is it for, and who should choose something else?
Span intends it for the prevention and treatment of pressure injuries; powered modes are for active wound treatment.
- The care team decides which surface suits the person's skin and risk.
- Do not use powered alternating pressure for an unstable spinal cord injury.
- Choose another surface for a user above 500 lbs (227 kg) or for low air loss.
Will it fit?
Match the width to the deck, then use the limit for the mode.
- 75 to 500 lbs (34 to 227 kg) non-powered; 400 lbs (181 kg) powered; the lower of that and the bed's limit applies.
- The 7 in (18 cm) height sets rail height. Health Canada advises gaps under 120 mm (4 3/4 in) in zones 1 to 3 and 60 mm (2 3/8 in) in zone 4.
- A used mattress cannot be returned under our return policy.
Using it day to day
Place it on the deck cover up; it self-inflates as a non-powered surface.
- With the 5900 digital unit, set the comfort level and choose FLOAT or ALTERNATE; AUTO FIRM firms the surface for 20 minutes for transfers or CPR.
- In a power failure it keeps a safe level; disconnect the air lines to reset it to non-powered.
- Wipe with neutral suds and lukewarm water, rinse and air dry for 20 to 30 minutes; avoid harsh cleaners and solvents.
What does it work with, and what are its limits?
It fits a standard hospital bed frame, and the control unit hangs on the footboard.
- The control unit is sold separately.
- Do not puncture the mattress; it voids the warranty.
- Do not move a patient on the mattress alone.
Which version should I choose?
This listing is the SC803629 mattress only.
- 5900 digital: FLOAT, ALTERNATE, AUTO FIRM, comfort levels and audible alarms, on a 10-minute cycle.
- 5500 standard: FLOAT and ALTERNATE with a comfort knob, on an 8-minute cycle, with no AUTO FIRM or audible alarm.
How does it compare?
The Custom Care Convertible LAL adds a low air loss cover and runs on the 8410 unit. The Easy Air XL is the bariatric option. The APM2 Safety Supreme is a powered alternating mattress without the foam shell. See hybrid mattresses.
Frequently asked questions
Does it work without the control unit?
Yes. It is non-powered reactive air, rated 75 to 500 lbs (34 to 227 kg).
What happens in a power failure?
It keeps a safe level of support; disconnect the air lines to reset it to non-powered.
Can the control unit stay connected?
No. Disconnect it when powered therapy ends so the mattress resets.
Funding may help: see who may help pay. Warranty terms are in the Warranty tab.
Specifications
- Weight capacity
- 75 to 500 lbs (34 to 227 kg) non-powered; 400 lbs (181 kg) powered
- Product weight
- 20 lbs (9 kg), mattress alone
- Construction
- High-density foam shell, Geo-Matt top with more than 800 cells, four interconnected air cylinders
- Cover
- LifeSpan bi-directional stretch with Shear Transfer Zones; bacteriostatic, fluid-proof, bleach resistant
- Cover care
- Wipe with neutral suds and lukewarm water, air dry 20 to 30 minutes; avoid harsh cleaners and solvents
- Therapy modes
- Non-powered reactive air; FLOAT, ALTERNATE and AUTO FIRM with a control unit
- Cycle time
- 10 minutes (5900 digital); 8 minutes (5500 standard)
- Comfort settings
- Adjustable comfort levels; AUTO FIRM for 20 minutes
- Alarms
- Low pressure and power failure, with audible alarm on the 5900
- CPR release
- AUTO FIRM plus a crash board
- Edge support
- Safety Edge foam bolsters
- Power failure protection
- Keeps a safe level; disconnect the air lines to reset to non-powered
- Power supply
- 100-120 V AC, 60 Hz (control unit)
- Pump weight
- 4.7 lbs (2.1 kg), control unit
- Mattress width
- 36" (91 cm)
- Mattress length
- 80" (203 cm)
- Mattress height
- 7" (18 cm)
- Warranty
- Mattress 5 years; cover 2 years; control unit 18 months, non-prorated
- Product type
- Air and Foam Mattress
Options and sizes can change some figures. Ask us to confirm yours before you order.
Documents
Delivery and service
Delivery and setup
We deliver and set up equipment across Toronto and the GTA, and show you how everything works. Custom and special-order products can take longer; we confirm timing with you before delivery.
Service and repairs
Our technicians service what we sell, at home or in our shop.
Try before you buy
Visit a showroom in Etobicoke or Brampton to try products first, or book a free home assessment.
Warranty
Warranty details
Span-America, part of Savaria Patient Care, warrants the PressureGuard Span-Care Convertible support surface and control unit to the original purchaser against defects in materials and workmanship, from the date of purchase. Repair or replacement is free during the term.
- Mattress: 5 years, not prorated.
- Cover: 2 years.
- Control unit: 18 months.
- Claims are made in writing with the original invoice and the sewn-in label or serial number; return shipping is paid by the maker with a return authorization number.
- Excludes defects from improper use, use without the cover, or use against the maker's instructions.
Replacement parts for this model